Repeatability, Precision

Repeatability-Blood
Interval Glucose concentration Coefficient of Variation (CV)
1 49.5 mg/dL 4.6%
2 78.1 mg/dL 3.2%
3 133.5 mg/dL 3.0%
4 200.7 mg/dL 2.9%
5 357.5 mg/dL 2.8%
Intermediate Precision-Control Solution
Interval Glucose concentration Coefficient of Variation (CV)
1 50.5 mg/dL 4.8%
2 117.5 mg/dL 2.7%
3 360.0 mg/dL 2.4%
 

FiggHealth smart blood glucometer

352 lay persons tested the capillary blood using the FiggHealth smart blood glucometer (y). The blood was taken from fingertip. Fingertip blood samples from the same subjects were also analyzed with YSI Model 2300 STAT PLUS Glucose Analyzer (x). The results were compared.

 
Linear Regression Results: FiggHealth Smart (y) vs. YSI Reference (x)
Sample Site Slope Intercept (mg/dL) R N
Fingertip 1.0134 -2.2871 0.9977 352
 

Fingertip blood samples were used for YSI reference measurement. The glucose concentration range was 23.3 to 516.0 mg/dL for FiggHealth Smart Blood glucometer testing with blood sample from fingertip test sites.

 
Fingertip Site: Consumers Accuracy Results
Within + 5% Within + 10% Within + 15% Within + 20%
71.9% (253/352) 96.0% (338/352) 99.4% (350/352) 100% (352/352)
 
Accurate Results
(Meter result is +/-15% of laboratory
result)
350 out of 352 (99.4% of results)
More Accurate Results
(Meter result is +/-10% of laboratory
result)
338 out of 352 (96.0% of results)
Most Accurate Results
(Meter result is +/-5% of laboratory
result)
253 out of 352 (71.9% of results)
 